대한핵의학회지 1998;32(5)425~8
원저 : Tungsten-188/Rhenium-188 발생기의 정도관리 ( Quality Control of Tungsten-188/Rhenium-188 Generator )
()
Author 장영수(Young Soo Chang),정재민(Jae Min Jeong),이동수(Dong Soo Lee),정준기(June Key Chung),이명철(Myung Chul Lee),
Affiliation
Abstract

Purpose: For the purpose of using Re-188 as a therapeutic radionuclide, we performed the quality control of the W-188/Re-188 generation system. Materials and Methods: Several quality control tests of the Re-188 eluate from generator were carried out of about 300 days. After elution of Re-188 with normal saline (20 ml), chromatogram and gamma-ray spectrum of Re-188 eluate were obtained. The presence of aluminum which was derived from the elumina bed of the generator was detected by using aluminum ion indicator kit. Re- 188 eluate was allowed to decay for several days, and then W-188 breakthrough in the Re-188 eluate was measured by detecting gamma-ray at 227 keV and 290 keV. The pH and the pyrogenicity of the eluate were checked. The Re-188 eluate from the generator was 67.4¡¾7.0% of W-188 during 270 days, and it was hightest at third day after previous elution. Radiochemical purity of Re- 188 eluate obtained from chromatogram was higher than 99%. Gamma-ray spectrum of Re-188 eluate showed a peak at 155 keV. Aluminum ion and W-188 contamination were not detected. The pH of Re-188 eluate was 3 and the concentration yield was 85%. Conclusion: Our experiments and results on quality control tests of Re- 188 eluate from W-188/Re-188 generator may be useful for setting W-188/Re-188 generator in hospitals.

Keyword W-188/Re-188 generator, Quality control, Radionuclide therapy
Full text Article 28502207.pdf 28502207.pdf
(03121)서울시 종로구 지봉로 29 금호팔레스빌딩 1705호
TEL : 02-745-2040 FAX : 02-745-3833 E-mail : ksnm@ksnm.or.kr
Copyright 2012 by The Korean Society of Nuclear Medicine(KSNM)