서로 다른 이중에너지 방사선흡수기계 기종 (Hologic QDR 4500-A 와 Lunar EXPERT-XL)간의 골밀도 교차 보정 ( Cross-calibration of Bone Mineral Density between Two Different Dual X-ray Absorptiometry Systems: Hologic QDR 4500-A and Lunar EXPERT-XL ) () |
Author |
조진만(Jin Man Jo),김재승(Jae Seung Kim),김기수(Ghi Su Kim),김상욱(Sang-Wook Kim),신중우(Jung Woo Shin),문대혁(Dae Hyuk Moon),이희경(Hee Kyung Lee), |
|
Affiliation |
|
Abstract |
Purpose: A cross-calibration equation is needed to
compare bone mineral density measured by different dual
X-ray absoptiometry systems. We performed this study to
establish appropriate cross-calibration equations
between two different dual X-ray absorptiometry
systems. Materials and Methods: Bone mineral density of
anterior-posterior lumbar spine (L2-4 level) and
femoral neck were measured in 109 women (55¡¾11yr)
using two different dual X-ray absorptiometry systems
(Lunar EXPERT-XL) and Hologic QDR 4500-A). Bone mineral
density values measured by two systems, including area,
bone mass content, bone mineral density and percentile
of young normals were compared and cross-calibration
equations between two systems derived. Results: The
bone mineral density values of 109 women measured by
Lunar system were 0.958¡¾0.17 g/cm2 at L2-4 and
0.768¡¾0.131 g/cm2 at femur neck, which were
significantly higher (13¡¾6% at L2-4 and 19¡¾7% at
femur neck, p<0.001) than those (0.851¡¾0.144 g/cm2
at L2-4 and 0.649¡¾0.108 g/cm2 at femur neck) by
Hologic system. Bone mineral content and percentile of
young normals measured by Lunar system were also
significantly higher than those by Hologic system
(p<0.001), whereas there was no difference in area
(p>0.05). There was a high correlation between bone
mineral density values of L2-4 and femoral neck
obtained with both dual X-ray absortiometry systems
(r=0.96 and 0.95, respectively). Cross-calibration
equations relating the bone mineral density were
Lunar=1.1287¡¿Hologic-0.0027 for L2-4 and
Lunar=l.1556¡¿Hologic+0.0l82 for femoral neck.
Conclusion: We obtained cross-calibration equations of
bone mineral density between Lunar EXPERT-XL and
Hologic QDR 4500-A. These equations can be useful in
comparing bone mineral density obtained by different
dual X-ray absorptiometry systems. |
Keyword |
Dual X-ray absorptiometry, Bone mineral density, Osteoporosis, Cross calibration |