대한핵의학회지 1999;33(3)282~7
서로 다른 이중에너지 방사선흡수기계 기종 (Hologic QDR 4500-A 와 Lunar EXPERT-XL)간의 골밀도 교차 보정 ( Cross-calibration of Bone Mineral Density between Two Different Dual X-ray Absorptiometry Systems: Hologic QDR 4500-A and Lunar EXPERT-XL )
()
Author 조진만(Jin Man Jo),김재승(Jae Seung Kim),김기수(Ghi Su Kim),김상욱(Sang-Wook Kim),신중우(Jung Woo Shin),문대혁(Dae Hyuk Moon),이희경(Hee Kyung Lee),
Affiliation
Abstract

Purpose: A cross-calibration equation is needed to compare bone mineral density measured by different dual X-ray absoptiometry systems. We performed this study to establish appropriate cross-calibration equations between two different dual X-ray absorptiometry systems. Materials and Methods: Bone mineral density of anterior-posterior lumbar spine (L2-4 level) and femoral neck were measured in 109 women (55¡¾11yr) using two different dual X-ray absorptiometry systems (Lunar EXPERT-XL) and Hologic QDR 4500-A). Bone mineral density values measured by two systems, including area, bone mass content, bone mineral density and percentile of young normals were compared and cross-calibration equations between two systems derived. Results: The bone mineral density values of 109 women measured by Lunar system were 0.958¡¾0.17 g/cm2 at L2-4 and 0.768¡¾0.131 g/cm2 at femur neck, which were significantly higher (13¡¾6% at L2-4 and 19¡¾7% at femur neck, p<0.001) than those (0.851¡¾0.144 g/cm2 at L2-4 and 0.649¡¾0.108 g/cm2 at femur neck) by Hologic system. Bone mineral content and percentile of young normals measured by Lunar system were also significantly higher than those by Hologic system (p<0.001), whereas there was no difference in area (p>0.05). There was a high correlation between bone mineral density values of L2-4 and femoral neck obtained with both dual X-ray absortiometry systems (r=0.96 and 0.95, respectively). Cross-calibration equations relating the bone mineral density were Lunar=1.1287¡¿Hologic-0.0027 for L2-4 and Lunar=l.1556¡¿Hologic+0.0l82 for femoral neck. Conclusion: We obtained cross-calibration equations of bone mineral density between Lunar EXPERT-XL and Hologic QDR 4500-A. These equations can be useful in comparing bone mineral density obtained by different dual X-ray absorptiometry systems.

Keyword Dual X-ray absorptiometry, Bone mineral density, Osteoporosis, Cross calibration
Full text Article 28502404.pdf 28502404.pdf
(03121)서울시 종로구 지봉로 29 금호팔레스빌딩 1705호
TEL : 02-745-2040 FAX : 02-745-3833 E-mail : ksnm@ksnm.or.kr
Copyright 2012 by The Korean Society of Nuclear Medicine(KSNM)