대한핵의학회지 (1967년~2009년)
대한핵의학회지 2004;38(1)74~84
소동물 Iodine-125 SPECT 개발을 위한 컴퓨터 시뮬레이션
(A Computer Simulation for Small Animal Iodine-125 SPECT Development)
Author 정진호, 최 용, 송태용, 정용현, 정명환, 홍기조, 민병준, 최연성, 이경한, 김병태,
Jin Ho Jung, Yong Choi, Ph.D., Yong Hyun Chung, M.S., Tae Yong Song, M.S., Myung Hwan Jeong, Key Jo Hong, Byung Jun Min, Yearn Seong Choe, Ph.D., Kyung-Han Lee,M.D. and Byung-Tae Kim, M.D.
Affiliation 성균관대학교 의과대학 생명의공학과, 삼성서울병원 핵의학과
Department of Biomedical Engineering, Sungkyunkwan University School of Medicine and Department of Nuclear Medicine, Samsung Medical Center, Seoul, Korea
Abstract

Since I-125 emits low energy (27-35 keV) radiation, thinner crystal and collimator could be employed and, hence, it is favorable to obtain high quality images. The purpose of this study was to derive the optimized parameters of I-125 SPECT using a new simulation tool, GATE (Geant4 Application for Tomographic Emission). Materials and Methods: To validate the simulation method, gamma camera developed by Weisenberger et al. was modeled. NaI(Tl) plate crystal was used and its thickness was determined by calculating detection efficiency. Spatial resolution and sensitivity curves were estimated by changing variable parameters for parallel-hole and pinhole collimator. Performances of I-125 SPECT equipped with the optimal collimator were also estimated. Results: In the validation study, simulations were found to agree well with experimental measurements in spatial resolution (4%) and sensitivity (3%). In order to acquire 98% gamma ray detection efficiency, NaI(Tl) thickness was determined to be 1 mm. Hole diameter (mm), length (mm) and shape were chosen to be 0.2:5:square and 0.5:10:hexagonal for high resolution (HR) and general purpose (GP) parallel-hole collimator, respectively. Hole diameter, channel height and acceptance angle of pinhole (PH) collimator were determined to be 0.25 mm, 0.1 mm and 90 degree. The spatial resolutions of reconstructed image of the I-125 SPECT employing HR:GP:PH were 1.2:1.7:0.8 mm. The sensitivities of HR:GP:PH were 39.7:71.9:5.5 cps/MBq. Conclusion: The optimal crystal and collimator parameters for I-125 imaging were derived by simulation using GATE. The results indicate that excellent resolution and sensitivity imaging is feasible using I-125 SPECT.

Keyword I-125, SPECT, collimator, simulation
Full text Article 3801074.pdf 3801074.pdf
(03121)서울시 종로구 지봉로 29 금호팔레스빌딩 1705호
TEL : 02-745-2040 FAX : 02-745-3833 E-mail : ksnm@ksnm.or.kr
Copyright 2012 by The Korean Society of Nuclear Medicine(KSNM)