소동물 영상을 위한 MicroPET R4 스캐너의 특성평가
(Performance Characteristics of MicroPET R4 Scanner for Small Animal Imaging)
Author 이병일.이재성.김진수.이동수.최창운1.임상무1,
Byeong Il Lee, Ph.D., Jae Sung Lee, Ph.D., Jin Su Kim, M.S., Dong Soo Lee, M.D.,Chang Un Choi, M.D.1, Sang-Moo Lim, M.D.1
Affiliation 서울대학교의과대학 핵의학교실, 원자력의학원1
Department of Nuclear Medicine, Seoul National University College of Medicine, Seoul, Korea, Korea Institute of Radiology & Medical Sciences, Seoul, Korea1
Abstract

Purpose: Dedicated animal PET is useful equipment for the study of new PET tracer. Recently, microPET R4 was installed in the Korea institute of radiology and medical science. In this study, we measured the characteristics of scanner. Materials and methods: Resolution was measured using a line source (F-18:65 μCi, inner diameter: 0.5 mm). The line source was put in the axial direction and was moved from the center of field of view to outside with 1 mm interval. PET images were reconstructed using a filtered back-projection and ordered subset expectation maximization. Line source (16.5 μCi, 78 mm) was put on the center of axial direction to measure the sensitivity when the deadtime was under 1%. Images were acquired during 4 minutes respectively from center to 39 mm outward. Delayed count was subtracted from total count and then decay was corrected for the calculation of sensitivity. Noise equivalent count ratio and scatter fraction were calculated using cylindrical phantom. Results: Spatial resolution of reconstructed image using filtered back-projection was 1.86 mm(radial), 1.95 mm(tangential), 1.95 mm(axial) in the center of field of view, and 2.54 mm, 2.8 mm, 1.61 mm in 2 cm away from the center respectively. Sensitivity was 2.36% at the center of transaxial field of view. Scatter fraction was 20%. Maximal noise equivalent count ratio was 66.4 kcps at 242 kBq/mL. Small animal images were acquired for confirmation of performance. Conclusion: Performance characteristics of microPET R4 were similar with reported value. So this will be a useful tool for small animal imaging.(Korean J Nucl Med 39(1):49-56, 2005)

Keyword MicroPET R4, Resolution, Sensitivity, Scatter fraction, Noise equivalent count ratio.
Full text Article 3901049.pdf 3901049.pdf
(03121)서울시 종로구 지봉로 29 금호팔레스빌딩 1705호
TEL : 02-745-2040 FAX : 02-745-3833 E-mail : ksnm@ksnm.or.kr
Copyright 2012 by The Korean Society of Nuclear Medicine(KSNM)