심장풀스캔에서 방사능 계측법에 의한 좌심실 용적의 측정 - 조직 감쇄의 보정에 의한 연구 - ( Left Ventricular Volume Measurement by Count Method with Attenuation Correction in Gated Blood Pool scan ) () |
Author |
범희승(Hee Seung Bom),정준기(June Key Chung),이명철(Myung Chul Lee),조보연(Bo Youn Cho),고창순(Chang Soon Koh),김지열(Ji Yeul Kim), |
|
Affiliation |
|
Abstract |
Attenuated end-diastolic and end-systolic left
ventricular counts which obtained from cardiac gated
blood poo1 scan were corrected using experimentally
calculated attenuation coefficient (μ=0.13/cm) and
depth of center of left ventricle. This method was
confirmed to be correct experimentally using phantom
balloon. To compare the accuracy of attenuated and
attenuation- corrected left ventricular volume
measurement authors studied 10 patients with ischemic
heart disease who underwent both gater blood pool sca
and X-ray contrast ventriculography within a week. The
attenuated and attenuation-corrected left ventricular
volume measured by count method correlated with
contrast ventriculographic volumes; however,
attenuation corrected measurement was correlated more
closely. |