폴라로그래피를 이용한 [201Tl]염화탈륨 주사액의 중금속 분석 ( Analysis of Heavy Metals in [201Tl]TlCl Injection Using Polarography )
()
Author 전권수(Kwon Soo Chun),서용섭(Yong Sup Suh),양승대(Seung Dae Yang),안순혁(Soon Hyuk Ahn),김상욱(Sang Wook Kim),최강혁(Kang Hyuk Choi),이동훈(Dong Hoon Lee),임상무(Sang Moo Lim),유국현(Kook Hyun Yu),
Affiliation
Abstract

Purpose: Thallous-201 chloride produced at Korea Cancer Center Hospital(KCCH) is used in detecting cardiovascular disease and cancer. Thallium impurity can cause emesis, catharsis and nausea, so the presence of thallium and other metal impurities should be determined. According to USP and KP, their amounts must be less than 2 ppm in thallium and 5 ppm in total. In this study, the detection method of trace amounts of metal impurities in [201Tl]TlCl injection with polarography was optimized without environmental contamination. Materials and Methods: For the detection of metal impurities, Osteryoung Square Wave Stripping Voltammetry method was used in Bio-Analytical System (BAS) 50W polarograph. The voltammetry was composed of Dropping Mercury Electrode (DME) as a working electrode, Ag/AgCl as a reference electrode and Pt wire as a counter electrode. Square wave stripping method, which makes use of formation and deformation of amalgam, was adopted to determine the metal impurities, and pH 7 phosphate buffer was used as supporting electrolyte. Results: T1, Cu and Pb in thallous-201 chloride solution were detected by scanning from 300 mV to -800 mV. Calibration curves were made by using TlNO3, CuSO4 and Pb(NO3)2 as standard solutions. Tl was confirmed at -450 mV peak potential and Cu at -50 mV. Less than 2 ppm of Tl and Cu was detected and Pb was not detected in KCCH-produced thallous-201 chloride injection. Conclusion: Detection limit of thallium and copper is approximately 50 ppb with this method. As a result of this experiment, thallium and other metal impurities in thallous-201 chloride injection, produced at Korea Cancer Center Hospital, are in the regulation of USP and KP. Polarograph could be applied for the determination of metal impurities in the quality control of radiopharmaceuticals conveniently without environmental contamination. (Korean J Nucl Med 2000;34:336-43)

Keyword Polarography, 201Tl, Osteryoung square wave stripping voltammetry, Quality assurance
Full text Article 28503272.pdf 28503272.pdf
(03121)서울시 종로구 지봉로 29 금호팔레스빌딩 1705호
TEL : 02-745-2040 FAX : 02-745-3833 E-mail : ksnm@ksnm.or.kr
Copyright 2012 by The Korean Society of Nuclear Medicine(KSNM)