변형 체내 표지법에 의한 적혈구 표지시 결합효율에 영향을 미치는 인자 평가 (The Evaluation of Factors Which Influence Binding Efficiency of Modified in Vivo Erythrocyte Labeling Technique) |
Author |
서한경1.김민우1.임석태1,2,3.손명희1,2,3, |
Han-Kyung Seo, M.S.1, Min-Woo Kim, M.D.1, Seok Tae Lim, M.D.1,2,3, and Myung-Hee Sohn, M.D.1,2,3 |
Affiliation |
전북대학교 의과대학 핵의학교실1, 의과학연구소2, 임상의학연구소3 Department of Nuclear Medicine,1 Institute for Medical Science,2 Research Institute of Clinical Medicine,3 Chonbuk National University Medical School, Chonju, Korea |
Abstract |
Purpose: We underwent this study to evaluate the factors which influence labeling efifciency when modified in
vivo erythrocyte labeling technique was used. Materials and methods: Thirty healthy volunteers (M:F=19:11, age:
25¡¾2 yrs) were enrolled in this study. Totally, two hundred ten samples were obtained from them. The 1 mg of
stannous pyrophosphate was injected intravenously at the beginning of labeling. After suitable tinning time (5 min,
20 min, 35 min) passed by, blood (5 mL, 3 mL or 1 mL) was withdrawn into 10 mL syringe previously containing
Tc-99m (740 MBq) and anticoagulant (heparin, ACD or CPDA) through 19-gauged scalp needle. The generator
ingrowth time of Tc-99m was within 24 hrs in each case. The blood samples were placed on rotating invertor
during incubation (10 min, 25 min, 40 min) but some of them were not. Immediately after the conclusion of
incubation, the labeled blood specimens to analyze were centrifuged. and then %Unbound Tc-99m was calculated.
Statical analysis was used paired T-test and one way ANOVA with SPSS 10.0. Results: The binding efficiency at 1
mL of blood volume was 73¡¾32%, 91¡¾10% at 3 mL and 96¡¾7% at 5 mL (p<0.01). The binding efficiency at 5 min
of tinning time was 45¡¾23%, 98¡¾6% at 20 min and 97¡¾8% at 35 min (p<0.001). The binding efficiency at 10 min
of incubation time was 96¡¾7%, 95¡¾12% at 25 min and 98¡¾3% at 40 min (p>0.05). The binding efficiency in case
of using rotating invertor was 96¡¾7% and the binding efficiency in case of not using it was 87¡¾18% (p>0.05).
There was no significant difference between them. In binding efficiency according to kinds of anticoagulants, ACD
was 98¡¾4%, CPDA was 97¡¾6% and heparin was 89¡¾20% (p<0.001). Conclusion: When modified in vivo
erythrocyte labeling technique is used with Tc-99m, the methods to obtain the highest labeling efficiency are as
follow. The withdrawing blood volume should be over 3 mL, tinning time should be kept between 20 min and 35
min, and incubation time should be kept between 10 min and 40 min. ACD or CPDA have to be used as a
anticoagulant except heparin and the blood samples should be placed on rotating invertor during incubation.
(Korean J Nucl Med 38(4):300-305, 2004) |
Keyword |
binding efficiency, modified in vivo method, erythrocyte labeling |