Tc-99m ECD 뇌혈류 SPECT 시행시 부적절한 영상획득 모수에 의한 인공산물 (An Artifact Caused by Using the Inadequate Acquisition Parameter in Tc-99m ECD Brain Perfusion SPECT) |
Author |
서한경1, 김민우1, 임석태1,2,3 손명희1,2,3, |
Han-Kyung Seo, M.S,1 Min-Woo Kim, M.D.,1 Seok Tae Lim, M.D.,1,2,3 and Myung-Hee Sohn, M.D.1,2,3 |
Affiliation |
전북대학교 의과대학 핵의학교실,1 의과학연구소,2 임상의학연구소3 Departments of Nuclear Medicine,1 Institute for Medical Sciences,2 and Research Institute of Clinical Medicine,3Chonbuk National University Medical School, Chonju, Korea |
Abstract |
Although artifacts in tomographic imaging can arise from a number of sources, this case presents an artifact
caused by using inadequate acquisition parameters which do not match the collimator with which a camera
equipped in Tc-99m ECD brain SPECT. Whenever the collimator in tomographic studies is changed, we should
take care to use suitable parameters of acquisition protocol according to the collimator with which a camera is
fitted. Other ways to prevent an error of this type is to use a point source.(Korean J Nucl Med 38(6):540-542, 2004) |
Keyword |
Artifact, SPECT, Brain, Tc-99m ECD |