Tc-99m ECD 뇌혈류 SPECT 시행시 부적절한 영상획득 모수에 의한 인공산물
(An Artifact Caused by Using the Inadequate Acquisition Parameter in Tc-99m ECD Brain Perfusion SPECT)
Author 서한경1, 김민우1, 임석태1,2,3 손명희1,2,3,
Han-Kyung Seo, M.S,1 Min-Woo Kim, M.D.,1 Seok Tae Lim, M.D.,1,2,3 and Myung-Hee Sohn, M.D.1,2,3
Affiliation 전북대학교 의과대학 핵의학교실,1 의과학연구소,2 임상의학연구소3
Departments of Nuclear Medicine,1 Institute for Medical Sciences,2 and Research Institute of Clinical Medicine,3Chonbuk National University Medical School, Chonju, Korea
Abstract

Although artifacts in tomographic imaging can arise from a number of sources, this case presents an artifact caused by using inadequate acquisition parameters which do not match the collimator with which a camera equipped in Tc-99m ECD brain SPECT. Whenever the collimator in tomographic studies is changed, we should take care to use suitable parameters of acquisition protocol according to the collimator with which a camera is fitted. Other ways to prevent an error of this type is to use a point source.(Korean J Nucl Med 38(6):540-542, 2004)

Keyword Artifact, SPECT, Brain, Tc-99m ECD