대한핵의학회지 (1967년~2009년)
Nucl Med Mol Imaging 2008;42(2)088~097
종설 : 핵의학 영상기기의 감마선 검출기
(Gamma-ray Detectors for Nuclear Medical Imaging Instruments)
Author 조규성,
Gyuseong Cho, Ph.D
Affiliation 한국과학기술원 원자력 및 양자공학과
Department of Nuclear and Quantum Engineering, KAIST, Daejeon, Korea
Abstract

In this review paper, basic configurations of gamma detectors in SPECT and PET systems were reviewed together with key performance parameters of the imaging system, such as the detection efficiency, the spatial resolution, the contrast resolution, and the data acquisition time for quick understanding of the system-component relationship and future design of advanced systems. Also key elements of SPECT and PET detectors, such as collimators, gamma detectors were discussed in conjunction with their current and future trend. Especially development trend of new scintillation crystals, innovative silicon-based photo-sensors and futuristic room- temperature semiconductor detectors were reviewed for researchers who are interested in the development of future nuclear medical imaging instruments. (Nucl Med Mol Imaging 2008;42(2):88-97)

Keyword SPECT, PET, collimator, scintillator, PM tube, APD, Si PM, CZT
Full text Article 4202088097.pdf 4202088097.pdf
(03121)서울시 종로구 지봉로 29 금호팔레스빌딩 1705호
TEL : 02-745-2040 FAX : 02-745-3833 E-mail : ksnm@ksnm.or.kr
Copyright 2012 by The Korean Society of Nuclear Medicine(KSNM)